Description
- Sensitive waterproof FHR transducer
- Flat surface design and waterproof TOCO transducer
- Fetal and maternal parameters
- Monitoring, storage and playback
- Three monitoring interfaces
- Integrated thermal printer
- Compatible with 150/152 mm universal recording paper
- Signals Overlap Verification to separate twins FHR
- Continuous post-delivery monitoring with maternal parameters: NIBP, SpO2, HR, ECG, TEMP
- Optional telemetry transducer system
- 24 hours waveforms playback
- Optional built-in wireless module
- Software for data transmission to PC
- 1, 2, 3 cm/minute real-time printing speeds
- Quick printing of stored waveform
- Rechargeable lithium-ion battery delivers 4 hours continuous monitoring
- Transducer mounting rack or wall mounting rack
- Handle for portability
Specifications:
- Basic Parameters: FHR, TOCO, AFM
- Internal Parameters: IUP, DECG (optional)
- Maternal Parameters: MECG, NIBP, MSpO2
- Display: 12.1″ (30.7 cm) colour TFT-LCD touchscreen
- Display Resolution: 800 x 600
- Dimensions: 347 mm x 330 mm x 126 mm
- Weight: 6 kg
- Battery: 14.8 V rechargeable lithium-Ion battery
- Ultrasound:
- Technique: Ultrasound Pulse Doppler with correlation
- Pulse Repetition Rate: 2 KHz
- Ultrasound Frequency: 1.0 MHz + 10%
- FHR Range: 50 bpm – 240 bpm
- Resolution: 1 bpm
- Accuracy: ±2 bpm
- DECG:
- Technique: Peak-to-peak detection
- FHR Range: 50 bpm – 210 bpm (International Standard)
- Resolution: 1 bpm
- Accuracy: ±1 bpm
- Input Impedance: > 10 M (Differential)
- Input Impedance: > 20 M (Common Mode)
- TOCO:
- TOCO Range: 0 – 100 Relative (%)
- Resolution: 1%
- Zero Mode: Auto/Manual
- Non-Linear Error: ±1 mmHg
- IUP:
- Pressure Range: 0 – 100 mmHg
- Resolution: 1%
- Zero Mode: Auto/Manual
- Non-Linear Error: ±mmHg
- Auto Fetal Movement (AFM):
- Technique: Pulsed Doppler Ultrasound
- Range: 0 – 100 (%)
- Resolution: 1%
- Marking: Manual fetal movement mark
- SpO2:
- Measuring Range: 0 – 100 Relative (%)
- Resolution: 1%
- Measuring Accuracy: ±2 bpm
- NIBP:
- Systolic Pressure: 50 mmHg – 240 mmHg
- Mean Pressure: 25 mmHg – 200 mmHg
- Diastolic Pressure: 15 mmHg – 180 mmHg
- Resolution: 1 mmHg
- Temperature:
- Measuring Range: 28°C – 44°C
- Accuracy: ±0.2°C